Skip to main content

Table 1 listing of physical parameters used for numerical simulation

From: Mathematical modeling of endovenous laser treatment (ELT)

 

Blood

Vessel wall

Perivenous tissue

References

μ a (mm-1) Deoxy-hemoglobin

0.20 (810 nm) 0.28 (980 nm)

0.1

0.030

[32,33]

μ' s (mm-1)

0.70 (810 nm) 0.6 (980 nm)

2.0

1.0

[32,33]

μ eff (mm-1)

0,86 (980 nm)

0.79

0.30

[32,34]

C (J.g-1.K-1)

3.82

3.78

3.78

[33,34]

ρ (g.mm-3)

1.05.10-3

1.05.10-3

1.05.10-3

[33,34]

k (W.mm-1.K-1)

5.6.10-4

5.6.10-4

5.6.10-4

[33,34]

Ea (J.mol-1)

4.48.105

4.30.105

4.30.105

[34,35]

A (s-1)

7.6.1066

5.6.1063

5.6.1063

[34,35]