Table 3 Sensitivity of ventricular fibrillation detection algorithms in per cent, wl = window length in seconds.

From: Reliability of old and new ventricular fibrillation detection algorithms for automated external defibrillators

Parameter wl Sns. if Spc. = 95 Sns. if Spc. = 99
TCI 3 15.0 1.0
TCI 8 25.3 1.3
ACF95 8 3.0 0.6
ACF99 8 3.0 0.6
VF 4 71.0 59.2
VF 8 73.4 59.7
SPEC 8 69.8 58.9
CPLX 8 38.8 5.8
STE 8 29.4 10.8
MEA 8 7.0 0.5
SCA 8 79.0 66.4
WVL1 8 56.7 35.2
LI 8 7.3 1.4
TOMP 8 9.1 1.8