Skip to main content

Table 5 Lesion Volume with 100% Normal Tissue Perfusion. Values represent the total volume of tissue necroses calculated over the course of the simulated ablation using various cell damage thresholds (D) and isothermal temperatures (IT).

From: Thermal modeling of lesion growth with radiofrequency ablation devices

Source Voltage (Volts)

D = 63% (mm3)

D = 100% (mm3)

IT = 42°C (mm3)1

IT = 47°C (mm3)1

IT = 60°C (mm3)

IT = 90°C (mm3)

2.5

0

0

0

0

0

0

5.0

0

0

0

0

0

0

7.5

0

0

0

0

0

0

10.0

0

0

1

0

0

0

12.5

0

0

47

0

0

0

15.0

0

0

185

6

0

0

17.5

0

0

358

68

0

0

20.0

4

2

513

177

1

0

22.5

23

19

784

287

15

0

25.0

131

89

1068

488

107

0

27.5

251

222

1502

759

241

3

30.0

433

364

2014

1064

423

24

  1. 1 – The 42°C and 47°C isothermal volumes were chosen specifically because they are frequently used to establish damage thresholds in hyperthermia and radiofrequency ablation, respectively.