Table 3 Comparison of data recorded in parallel with the emEEG system and a NicoletOne V32 system

From: Novel wireless electroencephalography system with a minimal preparation time for use in emergencies and prehospital care

Channel Examination room ICU Average
Subject 1 Subject 2 Subject 1 Subject 2
P9 1.3 ± 1.6 1.3 ± 1.6 1.0 ± 1.4 1.4 ± 1.9 1.3 ± 1.6
F7 1.1 ± 1.3 0.9 ± 1.0 0.8 ± 0.8 0.9 ± 0.9 0.9 ± 1.0
Fp1 1.2 ± 1.5 0.9 ± 1.2 0.7 ± 1.0 1.1 ± 1.9 1.0 ± 1.4
Fp2 1.0 ± 1.1 0.7 ± 0.8 0.7 ± 0.9 0.7 ± 1.8 0.8 ± 1.1
F8 1.3 ± 1.7 0.8 ± 1.1 0.7 ± 0.9 1.0 ± 1.0 0.9 ± 1.2
P10 1.2 ± 1.5 1.2 ± 1.5 0.9 ± 1.2 1.2 ± 1.7 1.1 ± 1.5
Average 1.2 ± 1.4 1.0 ± 1.2 0.8 ± 1.0 1.1 ± 1.6 1.0 ± 1.3
  1. The table displays the mean difference ± SD of the absolute difference signal computed from the EEGs recorded simultaneously with both systems (units: μV).