Table 1 Accuracy of the implemented v-ECG-based QRS peak detection algorithm during animal experiments

From: Effect of counter-pulsation control of a pulsatile left ventricular assist device on working load variations of the native heart

Animal (dur.) Occurred QRS peak Missed QRS peak Accuracy
1 (40 min) 3,957 39 99.0%
2 (60 min) 4,501 75 98.3%
3 (25 min) 1,826 61 96.7%
  1. dur time duration of peak number counting.