Skip to main content

Table 2 Test results to compare the SNR before and after filtering

From: Implementation of a portable device for real-time ECG signal analysis

Clean signal from

Type of

SNR before

SNR after

SNR

MIT-BIH Arrhythmia DB

noise

filtering (dB)

filtering (dB)

improvement (dB)

Record 118

BW

6

15.1689

9.1689

Record 118

EM

6

13.7682

7.7682

Record 118

MA

6

14.4875

8.4875

Record 118

BW

10

18.5482

8.5482

Record 118

EM

10

15.7853

5.7853

Record 118

MA

10

17.5477

7.5477

Record 118

BW

14

19.4872

5.4872

Record 118

EM

14

17.5643

3.5643

Record 118

MA

14

18.6872

4.6872

Record 119

BW

6

14.2548

8.2548

Record 119

EM

6

12.2346

6.2346

Record 119

MA

6

13.7458

7.7458

Record 119

BW

10

17.9658

7.9658

Record 119

EM

10

16.4867

6.4867

Record 119

MA

10

17.1054

7.1054

Record 119

BW

14

18.7548

4.7548

Record 119

EM

14

15.6875

1.6875

Record 119

MA

14

17.6561

3.6561

Average

 

10

16.3853

6.3853