From: Design of a microscopic electrical impedance tomography system for 3D continuous non-destructive monitoring of tissue culture

Figure 1

Design of the sample container included voltage sensing electrodes and current driving electrodes for the Kyung Hee University (KHU) Mark2 micro-EIT system. The EIT system injects current on surface electrodes and senses voltages on an additional set of surface electrodes. (a) The dimensions of a sample container and each 3 × 3 current driving electrodes on the current source plane and the current sink plane. (b) 24 × 15 voltage sensing electrodes on three imaging planes and dimensions for each sensing electrode. (c) Simulation of the voltage distributions subject to the primary current E 1 ± and (d and e) simulation results for two secondary currents, E 2 ± 1 , E 2 ± 2 .