Table 4 Data analysis results of each scanning condition

From: Application safety evaluation of the radio frequency identification tag under magnetic resonance imaging

RFID tag

Distance

Signal-to-noise ratio

Uniformity

Geometric distortion

High contrast resolution (lp)

SE

GRE

IR

SE

GRE

IR

SE

GRE

IR

SE

GRE

IR

With no tag

Group 0

85.002 ± 45.11

180.5214 ± 70.32

80.6595 ± 25.37

96.34%

96.80%

96.38%

0.0131 ± 8.27E-03

0.0124 ± 7.74E-03

0.0119 ± 5.97E-03

4

4

4

With tag

Group 1 (0 cm)

82.5572 ± 39.90

78.3693 ± 50.29

75.9142 ± 24.81

96.22%

97.66%

96.00%

0.0136 ± 1.09E-02

0.0127 ± 1.25E-02

0.0122 ± 1.15E-02

4

4

4

Group 2 (4 cm)

82.6249 ± 47.57

99.7621 ± 58.52

76.1715 ± 25.89

96.34%

97.22%

96.07%

0.0126 ± 8.73E-03

0.0131 ± 1.25E-02

0.0128 ± 9.73E-03

4

4

4

Group 3 (8 cm)

82.7500 ± 41.37

106.6630 ± 47.95

77.8441 ± 23.31

96.41%

97.03%

96.23%

0.0133 ± 8.59E-03

0.0129 ± 1.18E-02

0.0121 ± 9.82E-03

4

4

4

Group 4 (10 cm)

86.7404 ± 66.00

112.3880 ± 52.41

81.7903 ± 24.68

96.17%

96.87%

96.08%

0.0134 ± 1.09E-02

0.0113 ± 9.83E-03

0.0115 ± 9.72E-03

4

4

4

  1. Data analysis results of measured SNR, uniformity, geometric distortion and high-contrast resolution in each scan distance in the three scan sequences.