Table 1 The RFID Test library

From: Feasibility results of an electromagnetic compatibility test protocol to evaluate medical devices to radio frequency identification exposure

Frequency Occupied bandwidth Standard Program settings Maximum field strength
125 kHz (LF) Max ISO 14223 Type A Modulation Depth 100% 65 A/m
    ETU: 3 usec  
  Min ISO 14223 Type A Modulation Depth 90% 65 A/m
    ETU: 12 u sec  
13.56 MHz (HF) Max ISO 14443 Type A Pulse Width: 3 usec 12 A/m
    Transition Edge: 1 nsec  
  Min ISO 14443 Type A Pulse Width: 2 usec 12 A/m
    Transition Edge: 2.98 usec  
  Max ISO 14443 Type B Modulation Depth: 25% 12 A/m
    Transition Edge: 0.001 usec  
    SOF Low: 11 etu  
    SOF High: 2.5 etu  
    EOF Low: 10.5 etu  
  Min ISO 14443 Type B Modulation Depth: 15% 12 A/m
    Transition Edge: 1.17 usec  
    SOF Low: 10 etu  
    SOF High: 2 etu  
    EOF Low: 10 etu  
  Max ISO 15693 Modulation Depth: 100% 12 A/m
    Pulse Width: 9.44 usec  
    Transition Edge: 1 nsec  
  Min ISO 15693 Modulation Depth: 10% 12 A/m
    Pulse Width: 7.5 usec  
    Transition Edge: 800 nsec  
915 MHz (UHF) Max ISO 18000-6C Modulation: DSB-ASK 54 V/m
    Tari: 6.25 us  
    Pulse Width: 3.281 usec  
    Modulation Depth: 100%  
  Max ISO 18000-6C Modulation: DSB-ASK 54 V/m
    Tari: 25 us  
    Pulse Width: 0.265 Tari  
    Modulation Depth: 80%  
  Min ISO 18000-6C Modulation: PR-ASK 54 V/m
    Tari: 6.25 usec  
    Pulse Width: 1.656 usec  
  Min ISO 18000-6C Modulation: PR-ASK 54 V/m
    Tari: 25 usec  
    Pulse Width: 0.525 Tari  
2.45 GHz Max ISO 18000–4 Mode 1 Modulation Depth: 100% 54 V/m
    Transition Edge: 1 nsec  
    Tari: 25 usec  
  Min ISO 18000–4 Mode 2 Modulation Depth: 90% 54 V/m
    Transition Edge: 400 nsec  
    Tari: 33 usec  
  1. The RFID Test Library is a collection of test signals and field strengths that are representative for most commercially available RFID readers (RMS).