Table 5 Comparison of weaning rate between the study and control groups

From: Clinical Verification of A Clinical Decision Support System for Ventilator Weaning

 

Weaning Status

Total

 

Successful

Failure

 

Study group

103 (65.3%)

65 (38.7%)

168

Control group

75 (52.1%)

69 (47.9%)

144

Total

178

134

 
  1. Note: Not significant with a level of p<0.05 (odd ratio= 1.46; 95% CI=[0.929, 2.288]) using Pearson χ2 test.