Table 3 In-vivo and in-vitro measured angles

From: Characterization of the head-to-trunk orientation with handheld optical 3D apparatus based on the fringe projection technique

Range of Angle Mannequin Human subject ANOVA
mean ± std [°] mean ± std [°] F value/critical F
Rotation θ 51.48 ± 0.49 51.94 ± 0.89 3.1/4.2
Lateral tilt ψ 28.47 ± 0.73 29.05 ± 0.80 4.2/4.2
flexion/extension φ 40.67 ± 0.35 41.43 ± 0.88 20.7/4.2
  1. The number of repeated measurements is 15.