Probably not clinically significant | Device in operable | Probably significant | All experiments | ||
---|---|---|---|---|---|
Infusion Pump #1 (1995) | Occurrences of EMI | 0 | 4 | 0 | 4 (20%) |
Max Distance of EMI (cm) | - | 68 | - | 68 | |
RFID Reader Code causing EMI | N/A | 8,15,17,18 | N/A | 8,15,17,18 | |
Infusion Pump #2 (1996) | Occurrences of EMI | 1 | 4 | 0 | 5 (25%) |
Max Distance of EMI (cm) | 9 | 136 | - | 136 | |
RFID Reader Code causing EMI | 5 | 15-18 | N/A | 5,15-18 | |
Infusion Pump #3 (2002) | Occurrences of EMI | 1 | 0 | 0 | 1 (5%) |
Max Distance of EMI (cm) | 12 | - | - | 12 | |
RFID Reader Code causing EMI | 5 | N/A | N/A | 5 |