Skip to main content

Table 2 Impedances at some frequencies for the electrodes

From: Properties of screen printed electrocardiography smartware electrodes investigated in an electro-chemical cell

Electrode

|Z| at 1 Hz

|Z| at 15 Hz

|Z| at 50 Hz

|Z| at 2000 Hz

AAg

4200 (2200–6100)

1600 (1100–2200)

820 (600–1000)

240 (210–260)

AC

2200 (1200–3200)

1200 (1000–1300)

820 (800–850)

570 (520–620)

BAg

1880 (860–2800)

880 (760–1000)

560 (470–660)

260 (210–310)

BC

3400 (1900–5000)

1600 (1400–1700)

1000 (900–1100)

580 (530–630)

CAg

1300 (600–2100)

660 (490–830)

440 (350–520)

230 (190–270)

DAg

750 (530–980)

440 (340–550)

340 (260–420)

200 (170–230 )

Commercial

57 (47–68)

52 (43–60)

46 (38–53)

38 (32–43)

  1. The values are the mean and the 95% confidence interval (in brackets).