Table 5 Within-unit calibration accuracy for the H 3

From: A comparison of calibration data from full field digital mammography units for breast density measurements

Target/filter and kV combination

Within-unit Calibration: H3

120mAs

160mAsr

200mAs

Mo/Mo

   

kV =

25

49.4(0.53)

54.5(0.49)

56.1(0.48)

 

26

50.3(0.47)

53.7(0.40)

55.3(0.41)

 

27

51.6(0.36)

54.6(0.34)

55.1(0.35)

 

28

52.2(0.24)

54.6(0.23)

55.0(0.25)

 

29

53.4(0.16)

55.4(0.17)

55.6(0.19)

 

30

52.8(0.16)

55.1(0.20)

55.3(0.21)

 

31

53.3(0.27)

55.7(0.32)

55.7(0.32)

Mo/Rh

   

kV =

27

51.8(1.44)

53.6(1.52)

54.2(1.51)

 

28

52.6(1.45)

53.8(1.52)

54.0(1.53)

 

29

52.8(1.44)

54.0(1.52)

54.4(1.51)

 

30

53.7(1.35)

54.5(1.42)

54.8(1.42)

 

31

53.0(1.30)

54.6(1.37)

54.7(1.35)

 

32

53.3(1.16)

54.6(1.24)

54.2(1.19)

 

33

54.3(1.09)

55.0(1.15)

55.0(1.14)

 

34

55.6(0.94)

55.6(1.00)

55.8(0.98)

  1. Within-unit Calibration: 50/50 mixtures acquired and calibrated with the same FFDM unit.
  2. For each target/filter and kV setting, the 50/50 mixture phantom images were acquired with three mAs settings, including the reference denoted by mAsr. The mean and the standard deviation (SD) of the respective distributions are provided in PG units.