Figure 7 | BioMedical Engineering OnLine

Figure 7

From: Porous media properties of reticulated shape memory polymer foams and mock embolic coils for aneurysm treatment

Figure 7

Permeability of the SMP foams and MECs compared to that calculated for embolic coils [23]. The permeability of the MECs was higher than the permeability of both SMP foam types in all cases. Lowering the packing density resulted in an increase in permeability, while decreasing the hole density resulted in a decrease in permeability. The permeability data for embolic coils that was calculated using the Kozeny theory by Kakalis et al. [23] follows a similar trend as our measured MEC data.