Skip to main content

Table 1 Changes in HR, SpO 2 , and PWC-170 at 3,700 m (Mean ± SD, n = 8)

From: An oxygen enrichment device for lowlanders ascending to high altitude

 

Rest

Exercise

 

HR (beat/min)

SpO2(%)

PWC-170 (W)

Without device

83.00 ± 2.07

87.50 ± 1.51

201.26 ± 6.46

With device

75.00 ± 2.73*

95.75 ± 1.04*

230.43 ± 6.81*

  1. Note: *P<0.01. 'Without device’ vs. 'With device’. PWC-170 is the predicted power at HR = 170 beat/min, calculated from linear regression, see Materials and methods for detail.