Skip to main content

Table 3 Differences in mean ICP between simultaneous ICP signals

From: Simultaneous monitoring of static and dynamic intracranial pressure parameters from two separate sensors in patients with cerebral bleeds: comparison of findings

PatID Mean ICP (mmHg; average + std) Percentage of 6-s time windows with difference in mean ICP:
  Signal 1 Signal 2 > 5 mmHg > 10 mmHg
Category A
1 5.3 + 2.7 8.1 + 2.3 2 -
2 9.2 + 2.9 18.1 + 3.6 100 50
3 9.3 + 3.7 9.5 + 3.0 11 2
4 8.1 + 2.9 2.7 + 2.3 68 -
5 9.1 + 2.3 5.8 + 2.5 - -
Category B
6 15.3 + 3.4 20.3 + 2.3 37 2
7 9.3 + 4.6 7.6 + 5.5 35 4
8 6.9 + 4.1 9.5 + 5.3 37 -
9 −0.48 + 2.4 −29.6 + 4.6 100 100
10 11.9 + 5.4 12.5 + 5.2 5 -
Category C
11 11.4 + 2.6 14.8 + 2.3 4 3
12 9.3 + 3.5 9.5 + 3.4 1 1
13 10.0 + 3.6 9.0 + 5.1 19 6
14 10.3 + 2.8 5.9 + 6.5 14 9
15 7.8 + 14.3 8.5 + 4.1 27 6
16 14.1 + 1.9 10.6 + 3.7 14 1
17 15.8 + 5.5 15.5 + 3.8 8 5