Skip to main content

Table 2 CPAP Clinical Performance Results for 6 Subjects

From: Control system design for a continuous positive airway pressure ventilator

Subject
No.
Remark Weight
(kg)
Height
(m)
Age Sex Setting
Pressure
(cmH2O)
Max.
Pressure
(cmH2O)
Min.
Pressure
(cmH2O)
1 OSAS patient 101 1.72 43 Male 8.0 10.3 6.2
2 OSAS
patient
96 1.70 64 Male 7.0 8.6 4.6
3 Normal
subject
60 1.70 29 Male 8.0 9.6 6.2
4 Normal
Subject
60 1.78 29 Male 8.0 9.4 6.6
5 Normal
subject
58 1.69 26 Male 8.0 9.3 6.1
6 Normal
Subject
65 1.68 31 Male 8.0 10.1 6.8