From: Automatic detection of AutoPEEP during controlled mechanical ventilation
Parameters | True | N. of | Det. by SNT c | Det. by Sequential SNT c | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Id | Ventilator a | Lung model b | Label | breaths | P | N | Label | P | N | Label |
1 | PEP=0, Vt=500, f=15, P=0, I:E=1:2 | C=80, R=5 | N | 21 | 0 | 21 | N | 0 | 21 | N |
2 | PEP=0, Vt=500, f=15, P=0, I:E=1:2 | C=30, R=5 | N | 20 | 0 | 20 | N | 0 | 20 | N |
3 | PEP=0, Vt=500, f=25, P=0, I:E=1:2 | C=80, R=5 | P | 33 | 33 | 0 | P | 33 | 0 | P |
4 | PEP=0, Vt=500, f=25, P=0, I:E=1:1 | C=80, R=5 | P | 34 | 34 | 0 | P | 34 | 0 | P |
5 | PEP=0, Vt=300, f=20, P=0, I:E=1:2 | C=80, R=5 | N | 27 | 0 | 27 | N | 0 | 27 | N |
6 | PEP=0, Vt=500, f=12, P=0, I:E=1:2 | C=80, R=5 | N | 16 | 0 | 16 | N | 0 | 16 | N |
7 | PEP=0, Vt=500, f=20, P=15, I:E=1:3 | C=80, R=5 | N | 27 | 0 | 27 | N | 0 | 27 | N |
8 | PEP=5, Vt=500, f=20, P=0, I:E=1:3 | C=80, R=5 | N | 27 | 0 | 27 | N | 0 | 27 | N |
9 | PEP=5, Vt=500, f=20, P=0, I:E=1:2 | C=120, R=10 | P | 27 | 27 | 0 | P | 27 | 0 | P |
10 | PEP=0, Vt=700, f=20, P=0, I:E=1:2 | C=120, R=10 | P | 27 | 27 | 0 | P | 27 | 0 | P |
11 | PEP=0, Vt=700, f=20, P=0, I:E=1:6 | C=120, R=10 | P | 24 | 24 | 0 | P | 24 | 0 | P |
12 | PEP=0, Vt=700, f=20, P=0, I:E=1:1 | C=120, R=10 | P | 27 | 27 | 0 | P | 27 | 0 | P |
13 | PEP=0, Vt=700, f=20, P=0, I:E=1:2 | C=140, R=25 | P | 13 | 13 | 0 | P | 13 | 0 | P |