Table 4 Characteristics of the axial stiffness, reversible/irreversible deformation and peak load to failure following the static and dynamic loadings of the synthetic femur constructs with the implants LP, DCSP and ABP

From: Prospects of implant with locking plate in fixation of subtrochanteric fracture: experimental demonstration of its potential benefits on synthetic femur model with supportive hierarchical nonlinear hyperelastic finite element analysis

 

ABP (n = 3)

DCSP (n = 3)

LP (n = 3)

Axial stiffness (N/mm)

 

Mean ± SD

70.9 ± 16.3

110.2 ± 12.0

131.4 ± 10.8

Total Reversible Deformation (Displacement Amplitude, mm) at 400 N

Mean ± SD

12.4 ± 4.3

4.9 ± 1.4

4.1 ± 1.2

Total Irreversible/Plastic Deformation ((Displacement Amplitude, mm) at 1000 N

Mean ± SD

11.3 ± 0.8

2.4 ± 0.9

1.4 ± 0.2

Peak Cyclic Load to failure (N)

Mean ± SD

1100 ± 0

1167 ± 47.2

1600 ± 0

  1. The results were given as mean and standard deviation (SD).