Skip to main content

Table 1 Resolution to discern closely-spaced frequency components

From: Improved frequency resolution for characterization of complex fractionated atrial electrograms

Freq. (Hz)

Res-theor (Hz)

Res-NT (Hz)

Res-FT (Hz)

Significance

9.37 ± 0.47

0.19

0.21 ± 0.05

0.25 ± 0.08

P = 0.307

8.49 ± 0.24

0.15

0.23 ± 0.09

0.28 ± 0.08

P = 0.105

7.51 ± 0.32

0.11

0.18 ± 0.07

0.29 ± 0.18

P = 0.081

6.47 ± 0.27

0.09

0.14 ± 0.04

0.32 ± 0.20

P = 0.003

5.45 ± 0.32

0.06

0.13 ± 0.09

0.27 ± 0.18

P < 0.001

4.37 ± 0.27

0.04

0.08 ± 0.03

0.22 ± 0.07

P < 0.001

3.51 ± 0.29

0.03

0.07 ± 0.04

0.19 ± 0.06

P < 0.001

6.41 ± 2.01 (MN)

0.10

0.16 ± 0.10 (MN)

0.29 ± 0.22 (MN)

P < 0.001

  1. Freq.: the mean frequency used for the test, Res-theor – the theoretical resolving power of the new method, Res – NT: the resolving power of the new method, Res – FT: the resolving power of Fourier analysis, Significance – the significance of the difference between the resolving power of the techniques based on the unpaired t-test. MN – mean value.