Table 1 Comparison of mean respiratory rate during 5-minute measurements derived from the IRTR method and from conventional ECG measurement

From: Neonatal non-contact respiratory monitoring based on real-time infrared thermography

IRTR respiration derived parameters  
Subj RR (IRTR) [bpm] RR (ECG) [bpm] Minimum ROI temp [°C] Maximum ROI temp [°C]  
1 42.50 40.36 32.23 32.85  
2 44.25 42.60 32.11 32.77  
3 39.40 38.60 33.26 33.49  
4 45.14 45.20 32.18 32.45  
5 53.32 52.09 32.31 32.85