Skip to main content

Table 3 Values of the maximum electric field strength (Emax), minimum electric field strength (Emin) and electric field norm (EE) for each electrode configuration

From: Analytical and numerical solutions of the potential and electric field generated by different electrode arrays in a tumor tissue under electrotherapy

Types of electrode configurations

Emax

(V/mm)

Emin

(V/mm)

EE

(V/mm)

Configuration I (circle)

37.5855

0.0000

44.9545

Configuration II (ellipse)

37.8268

0.0002

39.7075

Configuration II-1 (ellipse)

8.3062

0.0001

17.6613

Configuration III (parabola)

3.2166

0.0000

10.0745

Configuration IV (hyperbola)

30.1479

0.0002

32.7382

Configuration IV-1 (hyperbola)

6.9787

0.0000

16.1356