Skip to main content

Table 3 Phantom volume measures and relative error for one-view, two-view, and three-view reconstructions

From: A 3D Freehand Ultrasound System for Multi-view Reconstructions from Sparse 2D Scanning Planes

Small Egg (7.2cc)

View

Number

View Locations

Frame

Number

Volume

Estimate

Relative

Error

 

1 view

Top window, short-axis

41

7.4218cc

3.08%

  

Top window, long-axis

44

5.1246cc

-28.83%

  

Side window, short-axis

47

7.4441cc

3.39%

  

Side window, long-axis

44

4.7494cc

-34.04%

 

2 views

Top window

Short-axis 21

Long-axis 22

7.4072cc

2.88%

  

Side window

Short-axis 24

Long-axis 22

7.3697cc

2.36%

 

3 views

Top & Side windows

Top window, short-axis 21

Top window, long-axis 22

Side window, long-axis 22

7.0045cc