Table 6 AED5 test results

From: Radiated radiofrequency immunity testing of automated external defibrillators - modifications of applicable standards are needed

Device designation

Type of interference observed

Field strength for worst-case failure (V/m)

Modulation

(Hz)

Frequency range of failure (MHz)

Device's behavior (Observations from user standpoint)

Notes

AED5

False negative

20 V/m

1 Hz pulsed (square wave)

38-41

and

47-49

Verbal warning heard during RF exposure while in analyze mode:

"Stop all motion",

"No one should touch the patient"

"shock not advised"

V-Fib applied - no shock advised.

No waveform distortion noted in the recorded waveforms.