From: Development and mechanical testing of a short intramedullary nail for fixation of femoral rotational osteotomy in cerebral palsy patients
Measure
Implant
Mean (SD)
Mean difference
CI 95%
LL
UL
p-value
Relative stiffness (N/mm)
Nail
268.26 (38.51)
-120.71
-154.84
-865.76
<0.001
Plate
388.97 (57.25)
Linear displacement (mm)
3.48 (0.43)
1.03
0.71
1.34
2.46 (0.44)