Skip to main content

Table 1 Comparative results for two electrodes with different radii.

From: Estimation of current density distribution under electrodes for external defibrillation

Electrode Radius [cm] Interfacing layer (5 mm thickness) Jmax/Jmin [Am-2] K=Jmax/Jmin
  ρ [Ωm] R [Ω ]   
5 20 1.25 1780/570 3.12
2.5 4 1.25 2450/800 3.06
5 80 5.1 1490/550 2.71
2.5 20 5.1 2150/780 2.76