Phantom-based experimental validation of fast virtual deployment of self-expandable stents for cerebral aneurysms

Abstract

Background

Endovascular intervention using a stent is a mainstream treatment for cerebral aneurysms. To assess the effect of intervention strategies on aneurysm hemodynamics, we have developed a fast virtual stenting (FVS) technique to simulate stent deployment in patient-specific aneurysms. However, quantitative validation of the FVS against experimental data has not been fully addressed. In this study, we performed in vitro analysis of a patient-specific model to illustrate the realism and usability of this novel FVS technique.

Methods

We selected a patient-specific aneurysm and reproduced it in a manufactured realistic aneurismal phantom. Three numerical simulation models of the aneurysm with an Enterprise stent were constructed. Three models were constructed to obtain the stented aneurysms: a physical phantom scanned by micro-CT, fast virtual stenting technique and finite element method. The flow in the three models was simulated using a computational fluid dynamics software package, and the hemodynamics parameters for the three models were calculated and analyzed.

Results

The computational hemodynamics in the patient-specific aneurysm of the three models resembled the very well. A qualitative comparison revealed high similarity in the wall shear stress, streamline, and velocity plane among the three different methods. Quantitative comparisons revealed that the difference ratios of the hemodynamic parameters were less than 10%, with the difference ratios for area average of wall shear stress in the aneurysm being very low.

Conclusions

In conclusion, the results of the computational hemodynamics indicate that FVS is suitable for evaluation of the hemodynamic factors that affect treatment outcomes.

Background

Cerebral aneurysms are pathological dilatations of cerebral vessels due to a weakening of the layers of the vessel wall. The rupture of cerebral aneurysms has catastrophic consequences, with high mortality and morbidity rates, as well as high health care costs. With the rapid improvement of medical imaging and neurovascular techniques, the treatment options for intracranial aneurysm are increasingly promising. The deployment of self-expandable stents typically plays a key role in aneurysm treatment options. Intracranial stenting is often used to provide a scaffold aimed at holding coils inside the aneurysm sac (stent-assisted coiling). It is assumed that such stents will also cause a reduction to the intra-aneurysmal flow, with the extent depending on the design of the stent [1]. However, aneurysmal interventions are complicated by the lack of knowledge on the specific conditions of each stent deployment.

In recent years, virtual stenting methods have been developed and applied to support clinicians in the planning of aneurysmal stenting procedures. To the best of our knowledge, the virtual stenting methods currently in use can be categorized into two types, according to the mechanism employed; these are the finite-element-method (FEM), and the fast virtual stenting (FVS) technique. Ding Ma et al. applied the finite element method to simulate the mechanical deployment of braided stents in patient-specific aneurysms [2]. Many other studies on virtual stenting using FEM have now been published [3–5]; these include qualitative and quantitative analysis, and FEM is now well accepted, although the computational time costs are high. The other published technique is FVS, which provides an estimation of the configuration of intracranial stents [6]. Ding Ma [7] proposed an enhanced fast virtual stenting method applied with a push–pull technique; this had the unique capability of accurately representing stent interventions in silicon models. However, quantitative validation of FVS analysis against experimental data has not been fully addressed. It is unclear if this method can reproduce the highly variable deployment configurations of patient-specific cases, and thereby enable accurate analysis of post-treatment hemodynamics. These analyses are vital to evaluate variations in treatment outcomes.

With the advances in endovascular treatments over the years, many studies on aneurysms have been performed, and many authors have reported on the hemodynamic factors related to aneurysms [8]. Knowledge of these hemodynamic characteristics is most certainly helpful for improving the understanding of the detailed procedures associated with aneurismal growth, evolution, and rupture. Computational fluid dynamics (CFD) is a robust non-invasive and promising tool that has been applied successfully to the study of hemodynamic changes in aneurysms. In order to illustrate the realism and usability of the novel FVS method, we performed in vitro studies of a patient-specific model with quantitative CFD analysis. We compared the experiments and simulations to check the quality of the numerical predictions, and to ensure they provided important information for clinical applications.

Methods

Patient specific aneurysm model

This study was approved by the ethics committee of Beijing Tiantan Hospital and informed consent was obtained from the study patient. A wide-neck internal carotid artery (ICA) aneurysm (width 12.27 mm, height 9.35 mm) was selected from digital subtraction angiography (DSA) image databases collected by our Interventional Neuroradiology Laboratory (Beijing Tiantan Hospital, China). The images of the patient-specific aneurysm were used to create a 3D reconstruction, which was then used to create a stereolithography format file of the initial geometry for the CFD simulation. The maximum diameters of the parent vessel and outlet branches at the trifurcation were 4.68, 3.01 mm (middle cerebral artery), and 2.64 mm (anterior cerebral artery).

A patient-specific aneurismal phantom was fabricated using stereolithography file and prototyping techniques. This model should reflect the real human anatomy, and was therefore considered to be well suited to our present study. Three views of the phantom with a deployed Enterprise stent are shown in Fig. 1: left lateral (LL), right lateral (RL), and posterior–anterior (PA).

Fig. 1
figure 1

Photographs of the patient-specific aneurismal phantom visualized by three views we defined. LL left lateral; RL right lateral; PA posterior–anterior

Deployment of the stent in the three simulated methods

The modeling workflow of the novel FVS deployment incorporated several simplifications [9]. First, the aneurysm geometry was isolated from the 3D patient-specific aneurism model using Geomagic Studio (Raindrop Geomagic, Research Triangle Park, NC, USA), and the centerline of the parent vessel was extracted using Mimics 10.01 (Materialise, Leuven, Belgium). After generation of the vessel centerline, a simplex mesh was initialized and treated as a deformable simplex model that was expanded using MATLAB 2013 (MathWorks, Natick, MA, USA). The deployment stopped when the simplex mesh achieved good apposition with the parent-vessel wall, thus giving the optimal deployed simplex mesh. The Enterprise stent patterns were then mapped onto the deployed simplex mesh and the wire was then swept, with the aim of obtaining the 3D structure of the stent.

In the patient-specific physical phantom model, an Enterprise stent was released into the parent artery of the saccular aneurysm by an experienced neurosurgeon. To conduct a CFD analysis of this experimental phantom, a micro-cone-beam computer tomography (micro-CT) system with a resolution of 10 µm was employed to scan the physical phantom and reconstruct the deployed Enterprise stent. The deployed stent was then reconstructed and a stereolithography format file was generated. An FEM deployment was also performed in accordance with the methods described in a previous study [10].

CFD simulation and hemodynamic analysis

Computational fluid dynamic analysis was performed for three scenarios: the stented aneurysm model from the realistic deployment, the FVS deployment, and the FEM deployment. The flow-governing Navier–Stokes equations were solved to second-order accuracy using ANSYS CFX 14.0 (ANSYS inc., Canonsburg, PA, USA). Velocity and pressure fields were computed with the common assumptions of incompressibility, Newtonian flow, rigid walls, and steady-state conditions. The density and viscosity of blood flow were input as 1050 kg/m3 and 4.0 cP respectively [11]. The mean flow rate at the inlet was set to maintain a wall shear stress of 1.5 Pa at the inlet, and the outlet pressure was set to 0 Pa. Several hemodynamic parameters, such as streamline, wall shear stress (WSS), area average of WSS on aneurysm, and average velocity inside the aneurysm, were calculated to validate the FVS according to CFD.

Results

As CFD models have been well validated, we used the CFD results for comparison purposes and to validate our FVS method. Qualitative visual comparisons between the FVS, FEM, and the realistic deployment of the Enterprise stent are shown in Fig. 2. The WSS, streamline, and velocity plane in the three different methods are all similar. Examination of WSS (Fig. 2a–c) shows that areas with distinct low WSS distributions are consistently found in the aneurysm, while areas with slightly higher WSS distributions are observed around the aneurysm neck. The velocity magnitudes within the aneurismal plane showed similar trends in all three models (Fig. 2d–f), with a larger area percentage of high velocity distributions in the FVS (Fig. 2e) and FEM (Fig. 2f) models. As was the case for the velocity magnitudes, the velocity streamlines of FVS and FEM were a little higher in the vortex than in the realistic phantom model (Fig. 2h–g).

Fig. 2
figure 2

Demonstration of wall shear stress a–c velocity plane, d–f and streamline, g–i in three stenting models created by different methods

For the quantitative comparison, the hemodynamic parameter, including area average of WSS on aneurysm, area average of WSS on parent vessel, and average velocity inside the aneurysm, were accurately calculated. The data are presented in Table 1 and the difference ratios of the parameters are presented in Table 2. We took the value of the hemodynamic parameters in the realistic phantom model as the reference, so that the difference ratios of the hemodynamic parameters between the phantom model stenting deployment and the FVS, and between the phantom model stenting deployment and the FEM, were calculated. All of the difference ratios were less than 10%, with the difference ratios for the area average of WSS on the aneurysm (Table 2) being very small. The average velocity within the aneurysm with the FVS method had a greater magnitude than with the realistic phantom, with the difference ratio between the two being approximately 10%. In contrast, the difference ratio for the average velocity within the aneurysm was very similar between the FVS and FEM methods. The area average of the WSS in the aneurysm (Pa) with the FVS method had a difference ratio of 1.26% from the realistic phantom, which was smaller than the 1.95% calculated between the FEM method and the realistic phantom. However, the difference ratio of the area average of WSS on the parent vessel (Pa) was slightly larger with FVS than with FEM (4.97 vs 2.83% respectively). There is no doubt that both our FVS method and the FEM method had differentiations with the method of realistic in some degree.

Table 1 Hemodynamic parameters of the parent vessel in the three different methods
Table 2 Percentage differences in the hemodynamic parameters of the parent vessel between the three different methods

Discussion

The present study evaluated the performance of the FVS method in stenting procedures for cerebral aneurysms. From a hemodynamic viewpoint, there were slight differences between the results of the FVS and those of the realistic phantom stenting deployment and FEM. As can be seen in Fig. 2, with the FVS procedure, the shape and geometry of the stent and the strut apposition to the vessel wall are distinctly different to the other two methods. Both the realistic phantom stenting deployment and the FEM showed malapposition to the parent-vessel wall, which may have been caused by the regions of curved vessel. It can also be observed that the shape of the stent using FEM was more similar to that of the realistic phantom. However, the difference ratios of the hemodynamic parameters calculated by CFD simulation are relatively small, so we therefore consider the FVS method to be acceptable. Additionally, we compared results against a physical phantom of the patient-specific cerebral aneurysm, and consider this to be an advantage of this research in comparison with aneurysm studies based only on idealized or simplified aneurysm models. The results of this study indicate that the combination of FVS with CFD should ultimately become a useful method for assessment of the best possible treatment options, and evaluation of the post-treatment prognosis.

The stent length, angle between the struts, and strut width, length, and diameter, should be taken into account, as these stent characteristics are sufficient to influence the hemodynamics and the treatment outcome. As is the case in most CFD simulations, we made the assumption of a rigid vessel wall, which allowed us to use Newtonian blood properties, and therefore, some variables such as vascular resistance, could not be quantitatively evaluated. Pulsatile blood flow, especially patient-specific flow velocity, could be applied to better quantify the accuracy of such simulations. There is no doubt that improved virtual modeling of stenting and coiling is urgently needed, as it can describe in a realistic manner the relevant hemodynamic factors and flow modifications. Improvements to FVS require more collaboration between engineers and clinicians.

Conclusions

In this work we performed numerical validation of FVS, a novel methodology for virtual stent release. The results indicate that FVS has the ability to perform virtual stenting, and that it will allow the evaluation of the flow dynamics and hemodynamics that would occur after the stenting of cerebral aneurysms. Accordingly, we are confident that our FVS technique provides an effective model of a stented aneurysm, and in combination with CFD, allows study of the ensuing hemodynamics. The method does suffer from some limitations; however, the results could still be validated.

In conclusion, the FVS technique, which is easier to automatically implement than FEM, may be a good approach to model the deployment of stents. If the clinicians are directly able to understand the FVS without the presence of an engineering expert, this approach should be strongly recommended for virtual stenting in endovascular interventional planning.

References

  1. Radaelli AG, Augsburger L, Cebral JR, et al. Reproducibility of haemodynamical simulations in a subject-specific stented aneurysm model–a report on the virtual intracranial stenting challenge 2007. J Biomech. 2008;41:2069–81.

    Article  Google Scholar 

  2. Ma D, Dargush GF, Natarajan SK, Levy EI, Siddiqui AH, Meng H. Computer modeling of deployment and mechanical expansion of neurovascular flow diverter in patient-specific intracranial aneurysms. J Biomech. 2012;45:2256–63.

    Article  Google Scholar 

  3. Bernardini A, Larrabide I, Morales HG, et al. Influence of different computational approaches for stent deployment on cerebral aneurysm haemodynamics. Interface Focus. 2011;1:338–48.

    Article  Google Scholar 

  4. Bernardini A, Larrabide I, Petrini L, et al. Deployment of self-expandable stents in aneurysmatic cerebral vessels: comparison of different computational approaches for interventional planning. Comput Methods Biomech Biomed Eng. 2012;15:303–11.

    Article  Google Scholar 

  5. Xiang J, Damiano RJ, Lin N, et al. High-fidelity virtual stenting: modeling of flow diverter deployment for hemodynamic characterization of complex intracranial aneurysms. J Neurosurg. 2015;123:832–40.

    Article  Google Scholar 

  6. Larrabide I, Kim M, Augsburger L, Villa-Uriol MC, Rufenacht D, Frangi AF. Fast virtual deployment of self-expandable stents: method and in vitro evaluation for intracranial aneurysmal stenting. Med Image Anal. 2012;16:721–30.

    Article  Google Scholar 

  7. Ma D, Dumont TM, Kosukegawa H, et al. High fidelity virtual stenting (HiFiVS) for intracranial aneurysm flow diversion: in vitro and in silico. Ann Biomed Eng. 2013;41:2143–56.

    Article  Google Scholar 

  8. Chung B, Cebral JR. CFD for evaluation and treatment planning of aneurysms: review of proposed clinical uses and their challenges. Ann Biomed Eng. 2015;43:122–38.

    Article  Google Scholar 

  9. Liu J, Fan J, Xiang J, Zhang Y, Yang X. Hemodynamic characteristics of large unruptured internal carotid artery aneurysms prior to rupture: a case control study. J Neurointerv Surg. 2016;8:367–72.

    Article  Google Scholar 

  10. Damiano RJ, Ma D, Xiang J, Siddiqui AH, Snyder KV, Meng H. Finite element modeling of endovascular coiling and flow diversion enables hemodynamic prediction of complex treatment strategies for intracranial aneurysm. J Biomech. 2015;48:3332–40.

    Article  Google Scholar 

  11. Ma D, Xiang J, Choi H, et al. Enhanced aneurysmal flow diversion using a dynamic push-pull technique: an experimental and modeling study. Am J Neuroradiol. 2014;35:1779–85.

    Article  Google Scholar 

Download references

Declarations

Authors’ contributions

QZ was responsible for CFD simulation and the preparation of the manuscript. ZM and KY developed FEM technique. YZ, LJ performed statistical analysis. YZ, JL, XY were responsible for the deployment of the realistic stent. Nikhil Paliwal and Hui Meng were responsible for virtual deployment of the stent. SW designed the research. All authors (1) have made substantial contributions to conception and design, or acquisition of data, or analysis and interpretation of data; (2) have been involved in drafting the manuscript or revising it critically for important intellectual content; and (3) have given final approval of the version to be published. Each author has participated sufficiently in the work to take public responsibility for appropriate portions of the content. All authors read and approved the final manuscript.

Acknowledgements

The research was supported by the National Natural Science Foundation of China (Grant Nos. 81571128, 81301003, 81471167, 81371315 and 81220108007), Special Research Project for Capital Health Development of China (Grant No. 2014-1-1071), and National Institutes of Health of the United States (Grant R01NS 091075).

Competing interests

The authors declare that they have no competing interests.

About this supplement

This article has been published as part of Bio Medical Engineering online volume 15 supplement 2, 2016. Computational and experimental methods for biological research: cardiovascular diseases and beyond. The full contents of the supplement are available online http://biomedical-engineering-online.biomedcentral.com/articles/supplements/volume-15-supplement-2.

Availability of data and materials

The data and materials are available upon request.

Ethics approval and consent to participate

This study was approved by the ethics committee of Beijing Tiantan Hospital and informed consent was obtained from the study patients.

Funding

Publication of this article was funded by the National Natural Science Foundation of China (Grant No. 81571128).

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Xinjian Yang or Shengzhang Wang.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, Q., Meng, Z., Zhang, Y. et al. Phantom-based experimental validation of fast virtual deployment of self-expandable stents for cerebral aneurysms. BioMed Eng OnLine 15 (Suppl 2), 125 (2016). https://doi.org/10.1186/s12938-016-0250-6

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/s12938-016-0250-6

Keywords